Zimmer GmbH Recall: Biolox Option Taper Sleeve, Type 1 Taper, -6mm Neck - Indicated for us
Zimmer GmbH Recall: Biolox Option Taper Sleeve, Type 1 Taper, -6mm Neck - Indicated for us is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Zimmer GmbH; Product: Biolox Option Taper Sleeve, Type 1 Taper, -6mm Neck - Indicated for use in total hip replacement with cemented or non-cemente
| Recalling firm | Zimmer GmbH |
|---|---|
| Product | Biolox Option Taper Sleeve, Type 1 Taper, -6mm Neck - Indicated for use in total hip replacement with cemented or non-cemented femoral and acetabular components. Item Number: 650-1064 |
| Reason for recall | Mislabeled: Incorrect sleeve in the packaging. The outer packaging is labeled as a standard neck sleeve, however, the product within the box is a -6mm neck sleeve, and vice versa. The neck of the sleeve can be identified by the device etching as either STD or -6 |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide distribution - US Nationwide and the countries of Japan, Netherlands. |
| Recall initiated | 20240808 |
| Recall number | Z-3133-2024 |
| Category | device |
| Country | Switzerland |